Predictive quality assurance of a linear accelerator based on the machine performance check application using statistical process control and ARIMA forecast modeling

Author(s)
Puyati, Wayo
Khawne, Amnach
Barnes, Michael
Zwan, Benjamin
Greer, Peter
Fuangrod, Todsaporn
Date Issued
August 1, 2020
Type
Article
DOI
10.1002/acm2.12917
Abstract
Purpose: A predictive linac quality assurance system based on the output of the Machine Performance Check (MPC) application was developed using statistical process control and autoregressive integrated moving average forecast modeling. The aim of this study is to demonstrate the feasibility of predictive quality assurance based on MPC tests that allow proactive preventative maintenance procedures to be carried out to better ensure optimal linac performance and minimize downtime. Method and Materials: Daily MPC data were acquired for a total of 490 measurements. The initial 85% of data were used in prediction model learning with the autoregressive integrated moving average technique and in calculating upper and lower control limits for statistical process control analysis. The remaining 15% of data were used in testing the accuracy of the predictions of the proposed system. Two types of prediction were studied, namely, one-step-ahead values for predicting the next day's quality assurance results and six-step-ahead values for predicting up to a week ahead. Results that fall within the upper and lower control limits indicate a normal stage of machine performance, while the tolerance, determined from AAPM TG-142, is the clinically required performance. The gap between the control limits and the clinical tolerances (as the warning stage) provides a window of opportunity for rectifying linac performance issues before they become clinically significant. The accuracy of the predictive model was tested using the root-mean-square error, absolute error, and average accuracy rate for all MPC test parameters. Results: The accuracy of the predictive model is considered high (average root-mean-square error and absolute error for all parameters of less than 0.05). The average accuracy rate for indicating the normal/warning stages was higher than 85.00%. Conclusion: Predictive quality assurance with the MPC will allow preventative maintenance, which could lead to improved linac performance and a reduction in unscheduled linac downtime.
Citation
Journal of Applied Clinical Medical Physics, 21(8), 73-82, 2020
Subjects

autoregressive integr...

machine performance c...

predictive quality as...

statistical process c...

Metrics
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your Institution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Accessibility settings
  • Privacy policy
  • End User Agreement
  • Send Feedback