Influence of Light Intensity on a Motion Artifact Signal in a Photoplethysmographic Signal

Author(s)
Koseeyaporn, Jeerasuda
Sinchai, Sakkarin
Tuwanut, Panwit
Wardkein, Paramote
Date Issued
November 27, 2018
Type
Conference Paper
DOI
10.1109/ICIIBMS.2018.8549938
Abstract
A photoplethygraphic (PPG) signal acquired from a commercial pulse oximeter is habitually disfigured by a motion artifact (MA) signal while any motion takes place. Inherently, the MA signal can either be additive or multiplicative, By having the MA signal intertwined with the PPG signal, the estimate of the oxygen saturation (SpO2) value is unreliable. The computation of SO02 value is based on proper red and infrared (IR) PPG signals measured by the pulse oximeter. Generally, the frequency components of the red and IR PPG signals as well as the MA signal are in the same range. To overwhelm this problem, the frequency components of the red and IR PPG signals are severed from the frequency band of the MA signal. To implement the separation, a legacy source of driving red and IR LEDs in the pulse oximeter is replaced by two alternating current sources having different frequencies. With this solution, the additive MA signal is no longer involved but the multiplicative MA signal is yet not concluded. In this work, change in light intensity fed to both red and IR LEDs is studied whether the changed light intensity has any impact on the multiplicative MA signal. According to the study, it is found that change in light intensity does not affect the multiplicative MA signal. The multiplicative MA signal still exists since the overall calculated SpO2 values have some error up to 1 %. Besides, change in light intensity does not improve the quality of the estimated SpO2 values while resting.
Citation
2018 International Conference on Intelligent Informatics and Biomedical Sciences Iciibms 2018, 8-14, 2018
Metrics
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your Institution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Accessibility settings
  • Privacy policy
  • End User Agreement
  • Send Feedback