Publication:
Finite element analysis of dental implant prosthetics

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Research Projects

Organizational Units

Journal Issue

Abstract

The paper presents the engineering analysis on stress characteristics of dental implant prosthetics. This research project focuses on a static simulation for dental implant using the three-dimensional finite element. Effects of four different Thread designs (Buttress Thread - US20140370459A1, Trapezoidal Thread - US20140212844A1, Reverse Buttress Thread - US20140816800A1 and a New Design) of dental implant prosthetics on stress are tested at four different areas (Abutment, Implant, Cortical Bone, and Cancellous Bone). Compressive forces, ranging from 60 to 200 N and Sheer force of 20 N with the force angle of 60°are applied to the dental implant models. Outcomes from this research project provide a better understanding of stress distribution characteristics that would be used for an improved design of dental implant Threads.

Description

Keywords

Compressive force, Dental implant prosthetics, Finite element, Shear force, Stress, Tread

Citation

Materials Today Proceedings, 5(3), 9525-9534, 2018