  4. Towards direct analysis of solid and liquid samples exploiting a 3D printed dialysis unit and sequential injection: Application for automated derivatization and determination of gamma-aminobutyric acid in foodstuff and beverages
Loading...
Thumbnail Image

Towards direct analysis of solid and liquid samples exploiting a 3D printed dialysis unit and sequential injection: Application for automated derivatization and determination of gamma-aminobutyric acid in foodstuff and beverages

Author(s)
Kasetsoontorn, Bhoonnarasa
Choengchan, Nathawut
Date Issued
February 8, 2020
Type
Article
DOI
10.1016/j.aca.2019.10.074
Abstract
In this work, a new design of a dialysis unit for direct analysis of solid and liquid samples is presented. The homemade unit was constructed using a 3D printer due to its simple and fast fabrication ability. The dialysis unit is composed of cylindrical-shaped donor and acceptor chambers. A stainless steel sieve is installed inside the donor chamber. SEM images clearly showed that the sieve prevented membrane blockage by suspension particles in the sample. Multiple dialysis units were connected to a sequential injection (SI) system for serial determination of gamma-aminobutyric acid (GABA) in solid and liquid samples. The dialysate from each dialysis unit was consecutively aspirated into the SI flow line for on-line derivatization of GABA with 2-hydroxy-1-naphthaldehyde (3.0% w/v). The derivative was detected spectrophotometrically at 425 nm. The linear calibration range extended to 1000 mg L−1 GABA (r2 > 0.99) with high precision (1.2 %RSD). The developed system was applied to analysis of dietary supplements, grains of germinated brown rice and milk. The samples were directly introduced into the donor chamber either as powder or liquids. The measured GABA content using the developed method was compared using high performance liquid chromatography, with good agreement using Pearson's correlation (r2 = 0.9999). The method has high accuracy based on recovery studies (99.8 ± 1.5%) and high sample throughput (64 samples h−1).
Citation
Analytica Chimica Acta, 1097, 103-109, 2020
Subjects

3D printing

Automated derivatizat...

Dialysis

Direct analysis

GABA

Sequential injection

Metrics
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your Institution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Accessibility settings
  • Privacy policy
  • End User Agreement
  • Send Feedback