Transparent Cross-Flow Platform as Chemiluminescence Detection Cell in Cross Injection Analysis
Author(s)
Somboonsuk, Thachkorn
Saetear, Phoonthawee
Mantim, Thitirat
Ratanawimarnwong, Nuanlaor
Wilairat, Prapin
Nacapricha, Duangjai
Date Issued
February 1, 2023
Type
Article
Abstract
This work presents the use of a transparent ‘Cross Injection Analysis’ (CIA) platform as a flow system for chemiluminescence (CL) measurements. The CL-CIA flow device incorporates introduction channels for samples and reagents, and the reaction and detection channels are in one acrylic unit. A photomultiplier tube placed above the reaction channel detects the emitted luminescence. The system was applied to the analysis of (i) Co(II) via the Co(II)-catalyzed H2O2-luminol reaction and (ii) paracetamol via its inhibitory effect on the catalytic activity of Fe(CN)63− on the H2O2-luminol reaction. A linear calibration was obtained for Co(II) in the range of 0.002 to 0.025 mg L−1 Co(II) (r2 = 0.9977) for the determination of Co(II) in water samples. The linear calibration obtained for the paracetamol was 10 to 200 mg L−1 (r2 = 0.9906) for the determination of pharmaceutical products. The sample throughput was 60 samples h−1. The precision was ≤4.2% RSD. The consumption of the samples and reagents was ca. 170 µL per analysis cycle.
Citation
Molecules, 28(3), 2023