Error analysis of tissue resistivity measurement
| dc.contributor.author | Tsai, Jang Zern | |
| dc.contributor.author | Will, James A. | |
| dc.contributor.author | Stelle, Scott Hubbard Van | |
| dc.contributor.author | Cao, Hong | |
| dc.contributor.author | Tungjitkusolmun, Supan | |
| dc.contributor.author | Choy, Young Bin | |
| dc.contributor.author | Haemmerich, Dieter | |
| dc.contributor.author | Vorperian, Vicken R. | |
| dc.contributor.author | Webster, John G. | |
| dc.date.accessioned | 2026-08-06T09:51:38Z | |
| dc.date.available | 2026-08-06T09:51:38Z | |
| dc.date.issued | 2002-04-30 | |
| dc.description.abstract | We identified the error sources in a system for measuring tissue resistivity at eight frequencies from 1 Hz to 1 MHz using the four-terminal method. We expressed the measured resistivity with an analytical formula containing all error terms. We conducted practical error measurements with in-vivo and bench-top experiments. We averaged errors at all frequencies for all measurements. The standard deviations of error of the quantization error of the 8-bit digital oscilloscope with voltage averaging, the nonideality of the circuit, the in-vivo motion artifact and electrical interference combined to yield an error of ±1.19%. The dimension error in measuring the syringe tube for measuring the reference saline resistivity added ±1.32% error. The estimation of the working probe constant by interpolating a set of probe constants measured in reference saline solutions added ±0.48% error. The difference in the current magnitudes used during the probe calibration and that during the tissue resistivity measurement caused ±0.14% error. Variation of the electrode spacing, alignment, and electrode surface property due to the insertion of electrodes into the tissue caused ±0.61% error. We combined the above errors to yield an overall standard deviation error of the measured tissue resistivity of ±1.96%. | |
| dc.identifier.citation | IEEE Transactions on Biomedical Engineering, 49(5), 484-494, 2002 | |
| dc.identifier.doi | 10.1109/10.995687 | |
| dc.identifier.issn | 00189294 | |
| dc.identifier.other | 2-s2.0-0036106121 | |
| dc.identifier.uri | https://dspace.kmitl.ac.th/handle/123456789/571 | |
| dc.source | IEEE Transactions on Biomedical Engineering | |
| dc.subject | Calibration | |
| dc.subject | Electrode | |
| dc.subject | Error analysis | |
| dc.subject | Error measurement | |
| dc.subject | Four terminals | |
| dc.subject | Noise measurement | |
| dc.subject | Probe constant | |
| dc.subject | Tissue resistivity | |
| dc.title | Error analysis of tissue resistivity measurement | |
| dc.type | Article |