Temperature measurement within myocardium during in vitro RF catheter ablation

dc.contributor.authorCao, Hong
dc.contributor.authorVorperian, Vicken R.
dc.contributor.authorTsai, Jang Zern
dc.contributor.authorTungjitkusolmun, Supan
dc.contributor.authorJe Woo, Eung
dc.contributor.authorWebster, John G.
dc.date.accessioned2026-08-06T09:50:43Z
dc.date.available2026-08-06T09:50:43Z
dc.date.issued2000-01-01
dc.description.abstractWhile most commercial ablation units and research systems can provide catheter tip temperature during ablation, they do not provide information about the temperature change inside the myocardium, which determines the lesion size. We present the details of a flow simulation and temperature measurement system, which allows the monitoring of the temperature change inside the myocardium during in vitro radio frequency (RF) cardiac catheter ablation at different blood flow rates to which the catheter site may be exposed. We set up a circulation system that simulated different blood flow rates of 0 to 5 L/min at 37°C. We continuously measured the temperature at the catheter tip using the built-in thermistor and inside the myocardium using a three-thermocouple probe. The system provides a means for further study of the temperature inside myocardium during RF catheter ablation under different flow conditions and at different penetration depths.
dc.identifier.citationIEEE Transactions on Biomedical Engineering, 47(11), 1518-1523, 2000
dc.identifier.doi10.1109/10.880104
dc.identifier.issn00189294
dc.identifier.other2-s2.0-0034330422
dc.identifier.urihttps://dspace.kmitl.ac.th/handle/123456789/319
dc.sourceIEEE Transactions on Biomedical Engineering
dc.subjectElectrical interference
dc.subjectRadiofrequency cardiac ablation
dc.subjectTemperature measurement
dc.subjectThermistor
dc.subjectThermocouple
dc.titleTemperature measurement within myocardium during in vitro RF catheter ablation
dc.typeArticle