Publication:
Design and Development of a Prototype for Measuring Range of Motion

Research Projects

Organizational Units

Journal Issue

Abstract

The ability to perform range of motion can help human maintain mobility and flexibility. The transitional device for measuring range of motion is Goniometer, a manual method. This method is depended on users' experience and skill which could lead to wrong interpretation and results. The purpose of this project is to design and develop a prototype for measuring Range of Motion (ROM) by using Inertial Measurement Units (IMUs). The prototype, called ROMie, is including hardware and software. ROMie had been validated for accuracy and sensitivity at 45, 60, 90 and 180 degrees. ROMie was tested for forty times at each angle,. The results showed that ROMie has the overall accuracy greater than 95 percent. Sensitivity is equal 1 at 95 percent confidence interval and greater than 0.875 at 99 percent confidence interval.

Description

Keywords

Accelerometer, gyroscope, IMU, range of Motion (ROM), wearable Sensor

Citation

IEEE International Conference on Industrial Engineering and Engineering Management, 2019-December, 1191-1195, 2018

Collections

Endorsement

Review

Supplemented By

Referenced By