Evaluation of rapid detection system for Escherichia coli in water samples

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

We demonstrated a fast detection platform for Escherichia coli (E. coli) in water. The system is based on a fluorescence detection technique by determining enzyme β-D-glucuronidase (GUD) activity and relating that activity to E. coli concentration. We found the portable E. coli detection platform is able to detect both the fluorescence 4-MU signal from the reaction of GUD assay from 0.01-1 μg/ml and to enumerate the number of E. coli in the range of 105-107 CFU/ml. The required incubation time is in between 20-240 min. In addition, we can estimate the maximum of incubation time for detecting minimum of E. coli 1 CFU/ml from the equation to be 576 mins. Our preliminary study demonstrates that the system can be used to enumerate the number of E. coli in water on-site and requires less detection time than traditional E. coli methods.

Description

Keywords

b-D-glucuronidase (GUD), drinking water, Escherichia coli, fluorescence, portable detector

Citation

Bmeicon 2016 9th Biomedical Engineering International Conference, 2017